SPECIFICATION FOR RO WATER

 
S.NO.                      TESTS                     SPECIFICATIONS
1. Description Clear, colorless and odorless liquid
2. pH Between 5.0 to 7.0
3. Total Hardness NMT 1 ppm
4. Conductivity NMT 30 µS/cm
5. Total Dissolved solids Nil
6. Free chlorine content Nil
7. Microbial Limit Test
a)Total Aerobic Microbial count NMT 100 cfu/ml
  b)Yeast and Moulds Nil
   

c)Specific pathogens

 

E. Coli Absent
Salmonella Absent
S. aureus Absent
P. aeruginosa Absent

 

Specification