SOP FOR PROCUREMENT, RECEIPT AND STORAGE OF PRIMARY STANDARD

SOP FOR PROCUREMENT, RECEIPT AND STORAGE OF PRIMARY STANDARD

1.0 OBJECTIVE

To lay down a standard procedure for procurements, receipt and storage of Primary Standard.

2.0 SCOPE

This procedure applies to all primary standards used for the standardization of the volumetric solutions and In-House Instrument                          Calibration in the (Name of Company)

3.0 RESPONSIBILITY

Officers / Executive – Quality Control shall be responsible for complies of this SOP

4.0  ACCOUNTABILITY

                   Manager- Quality Control and Head QA shall be accountable for compliance of this SOP.

5.0    ATTACHMENTS

         Format for label for primary standards                                   Attachment-I

Format for list of primary standards                                       Attachment-II

6.0  PROCEDURE

6.1    Commercially available AR/GR/Spectroscopy grade materials shall be procured and used for   preparation of the primary standards.

6.2    Indent shall have AR/GR/Spectroscopy grade mentioned specifically for primary standards.

6.3   On receipt of Primary Standards, seven digits Reference Code No. is allotted to each primary standard received as per the example given               below update the Primary Standards List.

Example : PS-001/11 where first two letter  stands for primary standards followed by (-) than 001 represent first primary standard received           followed by  “ / ” and 4 standards for the current year.

6.4    Take a clean and previously dried, sufficient number of narrow mouth and amber color screw capped vials for keep primary standards in                it. All the primary standards shall be treated as specified in the respective pharmacopoeia before storing it into the vials.

6.5    Each vial shall be labeled as per Format- SOP/QC/010/FT01-00.

6.6   Store the Primary Standards in desiccators unless otherwise mentioned on the label of the container.

6.7   If any specific storage conditions are mentioned on the label of the bottle, store the Primary Standards appropriately.

6.8   Label the desiccators containing Primary Standards, as “Primary Standards” and keep the updated Primary Standards List nearby.

6.9   Use the Primary Standards as per the “Directions of Use” mentioned on the label of the bottle. (e.g. Dry before use at 105°C for 4 Hr.)

9.10 Validity: Primary standards shall be valid up to two years after receipt or as per manufacturer’s specification.

7.0 REFERENCES

       In-House

8.0  ABBREVIATIONS

SOP: Standard Operating Procedure.

QC: Quality Control.

°C: Degree centigrade.

AR: Analytical Reagent.

e.g.: For example.

Hr.: Hour.

GR: General Reagent.

FT:Format.

PS:Primary standard

9.0 DISTRIBUTION LIST

Quality Assurance Department

Quality Control Department

10.0  HISTORY OF REVISION

Revision No. Date of Revision Reason for Revision

 

LIST OF PRIMARY STANDARDS

 

Sr. No. Name of Primary Standard Make Grade Mfg. LOT No. / Batch No. Reference Code  No.
SOPs